Product name:
Surgical mask (non-sterile)
Model:
Flat Type ear belt type
Intended use:
It is used for patients in order to reduce the risk of the spread of infections,and those are intended for use by healthcare professioonals in an operating room or in other medical settings with similar requirements.
Contraindication and warning:
None
Type of mask:
Type IIR
The performance characteristics of the device:
1. Bacterial filtration efficiency should be no less than 98% ;
2. The air exchange resistance between the two sides of the mask should less than 40 Pa/cm2.
3. Microbial cleanliness should not exceed 30 cfu/g.
Valid Date:
1 years.
Applicable technical standards:
EN 14683:2019+AC Medical face mask - Requirement and text methods.